» Ratings
30mA / 40 - 42kV / 0.3mAs
20mA / 43 - 63kV / 0.3-20mAs
15mA / 64 - 80kV / 0.3-20mAs*
* Max value = 13 mAs with DR system
» Generator
1.26kW output, inverter control
» kV setting
40-80kV (2kV step). Rotary switch for
digital display with 7 segment LED
» mAs selection
0.3-20mAs (32 settings)
Rotary switch for digital display with 7
segment LED
0.3 – 20 mAs range (19 settings)
» Total filtration
2.5mm AL. eq. at 80kV
» X-ray Tube
Toshiba brand
Focal spot: 0.8 x 0.8 mm
Heat unit : 7 KHU
Target angle : 16 degrees
» Power requirement
AC100 – 120V / 15A or AC200 – 240V / 8A
» Collimator
30 second timer and cross indication line.
» Line Cable and Hand Switch
Line cable : 6 meters long
Switch : Two stage with curled cable
» Ambient Conditions
Temperature : 20°C, +/- 15°C
Humidity : 69%, +/- 20%
Specifications are subject to Change without notice.